Preparations for a measuring run
7-19CenterMax Operating Instructions61211-1020202
Screw the metal plate on the measuring table (max. torque
40Nm).
Select the metal plate position so that probing of the probe sphere
is possible with all probes.
Place the calibration standard on the metal plate and turn the
rotary lever 90° to position I in order to activate the magnet.
The calibration standard can be placed in any position on the metal
plate.
Carrying out a reference measurement
The position of the calibration standard must be calibrated prior to
the length of the styli. Use the master probe for this purpose. Proceed
as follows:
1 Insert the master probe in the adapter plate receptacle
page 6-16.
2 Call the menu for reference measurement Operating instruc-
tions for the measuring software.
3 Select probe (stylus) 1 on the control panel.
Press the button between the five LEDs until the LED for probe
1 lights up.
4 Probe the sphere of the calibration standard to determine the
diameter of the sphere.
5 Remove the master probe from the adapter plate receptacle
Removing the probe (manually) on page 6-17.
Now, you can calibrate all probes.
Calibration procedure
After the reference measurement has been carried out, the probes
can be calibrated.
Semi-automatic calibration should be carried out for all probe sys-
tems.
Tensor calibration should be carried out for the VAST measuring
probe system. This is important since the measuring force causes a
1
3
2
4
5
NOTE
h
ss