Care
9-9CenterMax Operating Instructions61211-1020202
• Remove any particles or grease film from the probe elements and
the adapter plate.
• Replace any damaged styli.
Abrasion or material deposits
During scanning, the probe contacts the workpiece continuously. The
probe element glides along the surface of the workpiece. This may
cause abrasion and material deposits.
– Abrasion:
Material may be rubbed off the workpiece or probe element (e.g.
in case of sintered material) by the probe as it glides along the sur-
face of the workpiece during probing.
– Material deposits:
With certain materials (such as aluminum) material may be
removed from the workpiece and deposited on the probe element.
For this reason, probe elements should be checked regularly during all
scanning operations and replaced if necessary.
• Check the condition of the probe elements and clean them.
For information on how to remove aluminum deposits, please refer
to ➤ page 9-6.
• Replace the stylus if the probe element is damaged.
Cleaning/changing the filter mat
The interior of the control cabinet is cooled by a fan. The fan is
attached to the rear side of the control cabinet. The filter mat in the
intake hole must be cleaned every three months.
1Fan
2Air filter
3 Warning label: "Verschmutztes Luftfilter austauschen oder reinigen. /
Clean or replace soiled air filter.“
NOTE
1
3
2