Coordinate measuring machine
3-3CenterMax Operating Instructions61211-1020202
Connection data
Electric data
Compressed air
Certain requirements have to be met before connecting the com-
pressed air in order not to impair the function of the CMM.
Travel speed
Line voltage 100, 110, 115, 120, 125, 230,
240 VAC (±10%)
Type of current 1/N/PE
Frequency 50 to 60Hz (±3.5%)
Power consumption max. 2500VA
Fuse
for 230V or 240V C 25A
for 100V or 125V 25A
Requirements
Pressure 6 - 10bar
Degree of purity 99%; without solid particles, gas-
eous proportions of oil and water
Air consumption max. 20l/min with 5bar
Creep speed: SLOW function
0 to 5 mm/s
Setup mode: MAN operating mode
0 to 70mm/s
Series measuring operation: AUTO operating mode
Axis
Vector
max. 300mm/s