9-8
Maintenance and care
61211-1020202 CenterMax Operating Instructions
• Store the probe head in a dust-free and clean place, e.g. in a cabi-
net or drawer.
Cleaning:
Clean the probe head with a dry or damp cloth.
If the adapter plate receptacle is dirty:
• Clean the adapter plate receptacle with a mild cleaning agent.
• Then dry the adapter plate receptacle.
Make sure that the adapter plate receptacle is free from residues of
the cleaning agent. The function of the adapter plate receptacle may
be affected.
Probe rack
Vacuum the probe rack and the storage positions with a vacuum
cleaner and clean them with a mild cleaning agent.
• Remove all cleaning agent residues.
Concertina cover
Vacuum the concertina cover with a vacuum cleaner and clean it with
a damp cloth.
Important!
The cover must not be removed.
Housing covers
Vacuum the housing covers and clean them with a damp cloth.
Important!
The housing covers must not be removed.
Inspection measures
Checking the probe
Intensive care of the CMM also includes inspection of the compo-
nents used.
Regular inspection of the probes is required to ensure correct work-
piece measurement.
• Check the styli, other probe components and the adapter plate
regularly.
!
!