Overview
The components requiring care are listed in the following. The time
intervals apply only if all installation site requirements have been met.
Care measures
Measuring table
The workpieces must have full contact with the measuring table. Any
hard particle below the workpiece may lead to inaccurate measure-
ments. Furthermore, the surface of the measuring table and the
workpiece may be damaged.
Object Interval Type of care
Measuring table Every week or more fre-
quently
Vacuum cleaner,
cleaning agent
➤ page 9-5
Threaded holes in the
measuring table
Every week Vacuum cleaner ➤ page 9-6
Probe elements At least every week Depending on the degree
of soiling:
– cleaning agent.
– Special solvent; e.g.
KOH for removing alu-
minum deposits
➤ page 9-6
Calibration sphere Every week or more fre-
quently
Cleaning agent ➤ page 9-7
Adapter plate As required Fluff-free cloth; possibly
cleaning agent
➤ page 9-7
Probe head As required Dry or damp cloth ➤ page 9-7
Probe rack Every month Vacuum cleaner
clean with a cleaning
agent, if required
➤ page 9-8
Concertina cover Every month Vacuum cleaner ➤ page 9-8
Housing covers Every month Mild normal cleaning
agent
➤ page 9-8
Filter mat for ventilation
of the control cabinet
Every three months Clean by beating and vac-
uum with a vacuum
cleaner;
replace the filter, if
required
➤ page 9-9
Filter for compressed air As required Change ➤ page 9-10