Errors and faults
61211-1020202 CenterMax Operating Instructions
Errors occurring prior to the measuring run
If the measuring machine cannot be started, check whether the fol-
lowing errors can be excluded:
If the above mentioned errors can be excluded, put the CMM out
of operation and repeat the start-up procedure.
If it is then still not possible to carry out a measuring run, call a
ZEISS service engineer or our hotline page 8-9.
Error Cause Measure Check
The CMM cannot
be started.
Wrong starting
sequence.
Observe the specified sequence.
First, switch on the CMM and
then the computer.
Power supply Main switch is not
switched on.
Connect power supply:
Set main switch to position "1.
Position "1 must
be visible.
Control Control is not
switched on.
Switch on control:
Turn the rotary switch clock-
wise.
The indicator lamp
above the rotary
switch must light
up.
Drives The drives are not
switched on.
Switch on the drives:
Turn the rotary switch on the
control cabinet clockwise and
hold it for a few seconds.
The indicator lamp
above the switch
must light up.
EMERGENCY STOP
button on the con-
trol console is
locked.
Unlock the button:
Turn the EMERGENCY STOP
button slightly.
The button pops
out a few millime-
ters.
Quickstop button
on the control cabi-
net is locked.
Unlock the button:
Turn the quickstop button
slightly.
The button pops
out a few millime-
ters.
Pressure Compressed air line
defective.
Check the compressed air line
and the connections.
Indicating instru-
ments
Compressed air line
is not connected.
Connect compressed air line.