2-6
Description of the CMM
61211-1020202 CenterMax Operating Instructions
x direction
Crossbeam The x-bridge crossbeam is mounted on the columns and supports the
quill. The quill is guided on the crossbeam and can be moved back
and forth. Thus, probing in the x direction is possible.
y direction
Columns The drive and guideway side columns support the cross beam and the
quill and the frictionless air bearings enable movement of the com-
plete bridge in the y direction.
z direction
Quill The quill is guided along the crossbeam. It can be moved in the
vertical direction and allows probing in the z direction. The probe sys-
tem is located at the bottom of the quill.
Control
Control cabinet
32 bit / C99 The control for the CenterMax is located in a separate control cabinet.
The CenterMax is equipped with a 32 bit control. The control has the
designation C99.
IP54 The housing of the control cabinet has protection class IP54.
Front side of the control cabinet:
1 Operating elements
2 Type plate for control cabinet
3Door lock
4 Warning symbol for electric voltage: The control cabinet must be opened
only by an electrical engineer.
1
2
3
4