EasyManua.ls Logo

Zeiss CenterMax - Changing the Compressed Air Filters

Zeiss CenterMax
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9-10
Maintenance and care
61211-1020202 CenterMax Operating Instructions
Remove the air filter cover (2).
Clean or replace the filter mat.
Place the cover on the air filter.
Changing the compressed air filters
The compressed air unit is equipped with at least two filters. The
CMM is usually equipped with a fine filter and an extra fine filter. The
filters have to be checked regularly. The glass windows make it possi-
ble to check the degree of soiling.
Danger!
Switch off the compressed air before changing the filters.
A Fine filter
B Extra fine filter
!
AB
1
3
2
AB C
4
5
6

Table of Contents