2-16
Description of the CMM
61211-1020202 CenterMax Operating Instructions
Components
A VAST probe head
B Probe adapter plate and probe configuration
a
1 Adapter plate receptacle
2 Pin for positioning the probe.
limiting values
The limiting values for weight, length and torque of a probe must be
considered when assembling or mounting the probe.
Probe weight
The adapter plate receptacle may only be loaded with a maximum
weight of 600g.
Probe length
The length of the probe head including extension must not exceed
450mm.
A
2
1
B
600g
0
45
0
mm