Terminating a measuring run
7-33CenterMax Operating Instructions61211-1020202
Terminating a measuring run
Effect of system shutdown
If you put the CMM out of operation and switch it off completely, the
following data will be lost.
Reference point
Information on the coordinate systems.
When putting the machine into operation again, you must perform
another reference point travel.
Shutdown procedure
After the CMM is put out of operation, the probe head remains in its
last position. The space coordinates of its position are lost.
In order to avoid faults when the CMM is put back into operation, the
measuring run should be terminated correctly. For information on the
operators controls, please refer to page 2-8.
Proceed as follows:
CMM 1 Move the probe head to a safe position.
This position must be inside of the measuring volume. Prepare for
the next start-up by positioning the probe head close to the refer-
ence point.
Computer 2 Terminate the measuring software.
3 Power down the operating system.
4 Switch off the computer.
Control cabinet 5 Switch off the drives.
Press the quickstop button on the control cabinet and unlock it
again.
6 Switch off the control.
Set the rotary switch to the "OFF" position.
7 Switch off the power supply.
Turn the main switch counterclockwise to the "0" position.