9-4
Maintenance and care
61211-1020202 CenterMax Operating Instructions
Care
Although the CMM is robust and easy to care for, a certain degree of
care is still required. The sensitive measuring system of the CMM
requires a certain degree of cleanliness. Even the smallest dust parti-
cles on the probe element or the calibration sphere may cause inaccu-
rate measurements.
Cleaning and checking Measures for care also include the regular cleaning and checking of
certain components.
For cleaning, you require:
Mild detergent
A fluff-free cloth, e.g. made of linen
A vacuum cleaner
If necessary, special solvent for removing aluminum deposits from
the probe element, e.g. KOH (10%).
Danger!
When carrying out care measures, make sure that no travel move-
ments are performed.
Switch off the drives on the control cabinet in order to enable safe
execution of the care measures.
Important!
Do not use compressed air for cleaning.
Use fat dissolving and non-corroding cleaning agents. We recom-
mend Clean from JVA. This cleaning agent can be ordered from us
order no.: 84865.
Preventive care also includes making sure that all workpieces to be
measured are clean. The workpieces must be free from machining res-
idues (e.g. metal chips, oil) and dust.
Clean the workpieces before placing them on the measuring table.
!
!
!