Endocavity probes: Mindray recommends performing the Immersion Method of disinfection after
each patient exam. The use of a soft cloth thoroughly moistened with an approved disinfectant is
recommended for cleaning the strain relief, transducer housing and cable.
NOTE: The transducers cannot be sterilized. When sterility is required, use a sterile transducer
cover/sheath and sterile ultrasound/coupling transducer gel. Follow all manufacturer prescribed
techniques.
15.1.3 CLEANING & DISINFECTING
WIPE METHOD
1. Disconnect the transducer from the system. If present, remove the transducer
cover/sheath and discard.
2. Use a soft cloth dampened with mild soap (Liquinox) and water solution to
remove particulate, ultrasound coupling gel, or body fluids.
3. Examine the transducer for damage, such as cracks, holes, or fluid leaks. If damage is evident
discontinue use of the transducer and contact Mindray Technical Support.
4. Next, Mindray recommends the use of a disinfectant solution to wipe the
transducer housing, strain relief, and cable. Follow the manufacturer’s instructions for proper
cleaning.
Note: A list of approved disinfecting solutions can be found on page
5. Air dry (or towel dry) the transducer with a clean cloth.