Page 161 of 217
Error Code
Subsystem
Message for Error log
User Error Message
Action on Action
0xef100410
UCM
Software install failed. Please Re-
install the system software
"Software install failed.
Please Re-install the system
software."
Restart
0xef100412
UCM
Software license expired. Please Re-
install the system software.
"Software license expired.
Please Re-install the system
software."
Restart
0xef100412
UCM
Software license expired. Please Re-
install the system software.
"Software license expired.
Please Re-install "
"the system software."
Restart
0xef100413
UCM
Software license not authorized for
this release level. Contact system
support.
"Software license not
authorized for this release
level. Contact system
support."
Restart
0xef100414
UCM
Installed Software Release does not
support cart. Undock Scanner
"Installed Software Release
does not support cart.
Undock Scanner."
Restart
0xef104005
UCM
Failed to obtain device configuration
for DICOM image
"Internal System Error has
occurred. "
Restart
Battery Performance Charge Times - Reconditioning
The following information is intended to provide an overview on the battery-driven system operation
times and charging intervals for the Z-PAK battery pack.. There are a number of variables that impact
both the operating time and charge time, that must be taken into consideration when anticipating the
performance of a specific battery.
Both battery packs are "consumable" items, and as such should be expected to have a gradual drop-off
in performance (system operating time) over their life span. The system operating time numbers listed
in this document are in reference to a "new" (optimum) battery.
Charging times will be significantly impacted by the temperature of the battery pack at the time of the
start of the charging cycle. The intelligent PIC device that manages the charging process will disable,
or reduce the rate of charge, as dictated by the battery pack's current temperature.
The total time that will be required for a battery to reach a full state of charge is impacted by a number
of variables. These include:
1) Operating state of system (transducer in use: HV active)
2) Temperature of battery pack
3) Initial first-charge for Z-PAK:
Cart Battery Pack
Performance:
Typical system operation interval (on full charge): ....................................... up to 1.5 Hours