Validate that measured and calculated results shown in Calc Package reports reflect the clinical
observations.
Auto-Dop Trace is intended to serve as an adjunct to the diagnostic process in evaluating blood flow
during PW Doppler examinations. When using the Auto-Dop Trace feature, please evaluate the
results to verify that you are in agreement before committing the values to the Calc Report Package.
Always examine transducers for damage, such as cracks, splitting, holes, or fluid leaks. If damage is
evident, discontinue use of the transducer and contact Mindray.
Prior to initiating any disinfection process, disconnect the transducer from the ultrasound system.
Ensure that any connected external equipment, such as external monitors, printers and peripherals;
comply with relevant standards such as IEC60601-1 and IEC60601-1-2.
If an external video monitor is connected to the z.one
pro
system, it is necessary to ensure that an RF
ferrite is clamped to the cable as close to the z.one
pro
system as possible. Use a ferrite such as Fair-
Rite Products Corp. 0431167281 (or an equivalent).
The potential equalization terminal, located by the AC Mains connection, is connected to the system
chassis. It can be connected to corresponding terminals on other equipment to eliminate potential
differences. Do NOT use it for additional protective grounding
Damage to the system may cause poor image quality, including possible imaging artifacts sometimes
referred to as ‘halo’ or ‘headlight’ artifacts. Regularly inspect the system for damage and know how
to recognize imaging artifacts.
There are many types of system use and system error messages that might be displayed during the
use of the z.one
pro
Ultrasound System. If a message is encountered that is not self explanatory,
contact Mindray service for assistance.
Do not touch any of the connector contacts while performing a patient examination to prevent the
possibility of a hazardous current path.
Warnings - Ocular Imaging
To avoid injury to the patient, use only the Ocular Preset when imaging through the eye. The FDA
has established lower acoustic energy limits for ophthalmic use. The system will not exceed these
limits only if the Ocular Preset is selected
Warnings – Battery
To avoid electrical shock, do not touch the battery contact.
To avoid risk of fire, explosion, or burns:
Do not disassemble or alter the battery.
Do not short-circuit the battery by directly connecting the positive and negative terminals with
metal objects.
Do not heat or discard the battery in a fire.
Do not expose the battery to temperatures above 65° C (150° F).
Do not charge the battery near a heat source.
Do not leave the battery in direct sunlight.
Do not use a damaged battery.
Charge the battery at room temperature.
The battery should only be charged within the z.one
pro
System or a ZONARE provided battery
charger.