Page 154 of 217
Error Code
Subsystem
Message for Error log
User Error Message
Action on Action
0xaf050435
THERMALSER
VICES
XMT_V5NA over warning
Temperature Limit
"System Overheat Warning:
Finish exam and shutdown."
Continue
0xaf050436
THERMALSER
VICES
XMT_VSUPR over warning
Temperature Limit
"System Overheat Warning:
Finish exam and shutdown."
Continue
0xaf050438
THERMALSER
VICES
RCV_AFE over warning
Temperature Limit
"System Overheat Warning:
Finish exam and shutdown."
Continue
0xaf050439
THERMALSER
VICES
RCV_CW over warning Temperature
Limit
"System Overheat Warning:
Finish exam and shutdown."
Continue
0xaf05043a
THERMALSER
VICES
RCV_LV over warning Temperature
Limit
"System Overheat Warning:
Finish exam and shutdown."
Continue
0xaf05043b
THERMALSER
VICES
RCV_VSUPR over warning
Temperature Limit
"System Overheat Warning:
Finish exam and shutdown."
Continue
0xaf05043c
THERMALSER
VICES
Multiple Sensors over warning
Temperature Limit
"System Overheat Warning:
Finish exam and shutdown."
Continue
0xaf0f0404
ACQCONTRO
L
MPP can't connect
"MTP can't connect. Please
disconnect and reconnect."
Continue
0xaf0f0405
ACQCONTRO
L
MPP can't connect. Voltage Error
"MTP can't connect. Please
disconnect and reconnect."
Continue
0xaf0f0407
ACQCONTRO
L
CWBoard Current or Voltage Error
"CW Error. Please
disconnect and reconnect
MTP."
Continue
0xe0000401
SYSTEM_SER
VICES_EVA
Watchdog Server: Task Not
Responding
"Internal System Error has
occurred. "
Restart
0xe0004005
UNKNOWN
Failed to configure FPGAs);
"Internal System Error has
occurred. "
Restart
0xe0004005
UNKNOWN
Fatal Error due to NVASSERT. See
above NVASSERT log statement for
details
"Internal System Error has
occurred. "
Restart
0xe00a0401
IMAGING
PIC is Invalid, No HV support
"Internal System Error has
occurred. "
Restart
0xe00a0403
IMAGING
HV's not the same when attempting
dynamic update of limits
"Internal System Error has
occurred. "
Restart
0xe00a0404
IMAGING
Failed to reset the PIC HV task
"Internal System Error has
occurred. "
Restart