XMT_V5NA over warning
Temperature Limit
"System Overheat Warning:
Finish exam and shutdown."
XMT_VSUPR over warning
Temperature Limit
"System Overheat Warning:
Finish exam and shutdown."
RCV_AFE over warning
Temperature Limit
"System Overheat Warning:
Finish exam and shutdown."
RCV_CW over warning Temperature
Limit
"System Overheat Warning:
Finish exam and shutdown."
RCV_LV over warning Temperature
Limit
"System Overheat Warning:
Finish exam and shutdown."
RCV_VSUPR over warning
Temperature Limit
"System Overheat Warning:
Finish exam and shutdown."
Multiple Sensors over warning
Temperature Limit
"System Overheat Warning:
Finish exam and shutdown."
"MTP can't connect. Please
disconnect and reconnect."
MPP can't connect. Voltage Error
"MTP can't connect. Please
disconnect and reconnect."
CWBoard Current or Voltage Error
"CW Error. Please
disconnect and reconnect
MTP."
Watchdog Server: Task Not
Responding
"Internal System Error has
occurred. "
Failed to configure FPGAs);
"Internal System Error has
occurred. "
Fatal Error due to NVASSERT. See
above NVASSERT log statement for
details
"Internal System Error has
occurred. "
PIC is Invalid, No HV support
"Internal System Error has
occurred. "
HV's not the same when attempting
dynamic update of limits
"Internal System Error has
occurred. "
Failed to reset the PIC HV task
"Internal System Error has
occurred. "