IEC 60601-1-2: 2007 Certification
The z.one
pro
system is suitable for use in all establishments other
than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
IEC 61000-3-2 harmonic emissions
IEC 61000-3-3 voltage fluctuations/flicker
emissions
IEC 61000-4-2 electrostatic discharge
(ESD)
Compliance level:
±6 kV contact
±8 kV air
Electromagnetic environment guidance:
Floors should be wood, concrete, or ceramic tile. If floors are
covered with a synthetic material, the relative humidity should be
at least 30%.
IEC 61000-4-3 Radiated Field Immunity
80MHz - 2.5GHz
3V/m, 80%@2Hz
IEC 61000-4-4 electrical fast transients
(EFT)
±2 kV for AC Mains
±1 kV for I/O lines
Electromagnetic environment guidance:
Main power quality should be that of a typical commercial or
hospital environment.
±1 kV differential mode
±2 kV common mode
Electromagnetic environment guidance:
Main power quality should be that of a typical commercial or
hospital environment.
IEC 61000-4-6 Conducted Immunity
0.15MHz - 80 MHz
3Vrms @ 2Hz
IEC 61000-4-8 power frequency (50/60
Hz) magnetic field immunity
3A/m, 50/60Hz
Electromagnetic environment guidance:
Power frequency magnetic fields should be at levels characteristic
of a typical commercial or hospital environment.
IEC 61000-4-11 voltage dips, short
interruptions, and voltage variations on
power supply input lines
Electromagnetic environment guidance:
Main power quality should be that of a typical commercial or
hospital environment.