4. To save the new settings, select APPLY from the main menu, and press one of the SET keys to
save new settings.
DICOM” MPPS” (MODALITY PERFORMED PROCEDURE STEP) CONFIGURATION
1. Within the DICOM selection, use the Menu Control to backlight MPPS (Modality Performed
Procedure Step) selection, press Select button to bring up this menu.
DICOM Sub-Menu DICOM MPPS Sub-Menu
2. To add a new MPPS destination, select the NEW from the main menu, to bring up the following sub-
menu.
3. Using the alpha-numeric keyboard, enter in the DICOM configuration values/setting in the appropriate
fields, as specified in the information previously provided in the “Pre-Installation Site Survey Form”,
by the medical facility’s IT System Administrator.
4. To save the new settings, select APPLY from the main menu, and press one of the SET keys to
save new settings.