3-6
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Principles of Operation
Sample Analysis Cycle Overview
Section 3
Results Displayed
All data is transmitted to the Data Module for analysis. Results are computed for
all parameters and are displayed on the RUN screen. Results are also stored in a
log format called the Data Log.
Instrument Flushed
1. The remaining sample segment from the aspiration process is flushed to
Waste Chamber #2.
2. The remaining segments in the WBC and RBC Mixing Chambers are flushed
to Waste Chamber #3.
3. The segments sent to the Optical Flow Cell are flushed to Waste Chamber #1.
Instrument Rinsed
1. The Open Sample Aspiration Probe is rinsed internally and externally with
diluent/sheath.
2. In Closed Mode, the needle is rinsed internally and externally with diluent/
sheath.
3. The WBC and RBC/PLT Mixing Chambers are rinsed with diluent/sheath.
4. The Optical Flow Cell and Sample Line tubing are rinsed with diluent/sheath.
5. The HGB Flow Cell is rinsed with diluent/sheath.