CELL-DYN
®
3200 System Operator’s Manual
7-1
9140181K—July 2002
Section 7 Operational Precautions and Limitations
S
ect
i
on 7
O
perat
i
ona
l
Precaut
i
ons an
d
L
i
m
i
tat
i
ons
Overview
This section deals with the precautions that need to be taken to ensure the
performance and validity of the CELL-DYN 3200 System and its test results.
The following system precautions and limitations are reviewed in this section:
General Limitations
Location Requirements
Reagent Storage and Handling
Printer Precautions
Limitations
The CELL-DYN
3200 System is designed for in vitro diagnostic use.
Abbott has designed the CELL-DYN 3200 System components for optimal
performance. Substituting reagents, calibrators, controls, and components
manufactured by other companies may adversely affect the performance of
the instrument.
Follow the recommended maintenance schedules and procedures as outlined
in Section 9: Service and Maintenance.
During the warranty period, all service and repair must be performed by
Abbott-authorized representatives.
Location Requirements
The location of the CELL-DYN 3200 System is an important consideration that
affects proper instrument functioning, operating safety, and ease of use.
An Abbott-authorized representative must install the instrument.
The location should have nonporous, nonabsorbing work surfaces and
flooring that can be cleaned easily and disinfected using recommended
procedures.
Place the CELL-DYN 3200 System on a hard, level surface. Locate the
system:
away from direct sunlight.
away from the path of a cooled air or heated air outlet.
away from other instruments that may interfere with the
CELL-DYN 3200 System, such as drying ovens, centrifuges, x-ray
equipment, MRI (magnetic resonance imaging) equipment, CRTs or
computers, video terminals, copiers, ultrasonic cleaners, and patient
areas. Please note that the CELL-DYN 3200 System has been evaluated
to EN55011 and EN61000 for electromagnetic emissions and immunity,
respectively.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott CELL-DYN 3200 and is the answer not in the manual?

Abbott CELL-DYN 3200 Specifications

General IconGeneral
BrandAbbott
ModelCELL-DYN 3200
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals