CELL-DYN
®
3200 System Operator’s Manual
2-17
9140181H—October 2001
Section 2 Installation Procedures and Special Requirements
Power On
1. Turn the instrument’s Main Power Switch to ON. When the initialization
process is complete, press the [RUN] key to prime the instrument. When the
priming process is complete, the READY light on the Analyzers indicator
panel is lit and the message READY is displayed in the Status Box on the
monitor.
2. Prime the Sample Loader by running several blood samples in the Closed
Mode. If necessary, refer to the directions given in Section 5: Operating
Instructions; Subsection: Sample Analysis on the CELL-DYN 3200SL,
Running Samples, Closed Mode Procedure.
3. Confirm that the background count is acceptable before running patient
samples.