CELL-DYN
®
3200 System Operator’s Manual
2-5
9140181K—July 2002
Section 2 Installation Procedures and Special Requirements
Tubing Installation
Reagent and Waste Tubing
1. Locate the reagent tubing in the Accessory Kit.
2. Inspect each length of tubing carefully for damage or cracks.
3. Attach the non-weighted end of the tubing with the Purple WBC Lyse label
to the Purple Connector on the right side of the Rear Panel. (Refer to Figure
2.1.) This reagent is for the WBC dilution. Wipe the outside of the tubing
with a damp lint-free pad (such as DYN-A-WIPE) and place the weighted
end into the container of CELL-DYN WBC Lyse. Secure the cap. Place the
container on the same level as or lower than the unit.
4. Attach the non-weighted end of the tubing with the Red Diluent/Sheath
label to the Red Connector. This reagent is for the RBC dilution and for
rinsing the flow system. Wipe the outside of the tubing with a damp lint-free
pad (such as DYN-A-WIPE) and place the weighted end into the container of
CELL-DYN Diluent/Sheath. Secure the cap. Place the container on the same
level as or lower than the unit.
5. Attach the non-weighted end of the tubing with the Blue HGB Lyse label to
the Blue Connector. This reagent is for the HGB dilution. Wipe the outside
of the tubing with a damp lint-free pad (such as DYN-A-WIPE) and place the
weighted end into the container of CELL-DYN HGB Lyse. Secure the cap.
Place the container on the same level as or lower than the unit.
6. Attach the Waste Outlet Tubing to the :DVWH Outlet Connector. Place the
end of the tubing with the cap and sensor into the waste collection container.
Ensure that the waste collection container is adequately labeled. Secure the
cap, or remove the cap from the tubing and place the tubing into a drain
suitable for collection of waste with possible biological and chemical hazard.
Be sure that the tubing is secured to the drain hole.