CELL-DYN
®
3200 System Operator’s Manual
4-15
9140181H—October 2001
Section 4 Performance Characteristics and Specifications
Performance Characteristics
Typical Precision
Within Sample
The pooled precision values (CVs) for the Hemogram parameters, shown in
Table 4.14, are based on the analysis of data from N = 20 replicate runs based on
the same sample.
NOTE: N=20 is based on one whole blood sample aliquotted into 4 tubes, with
each tube run 5 times.
The data was obtained from several CELL-DYN 3200 systems over a period of
weeks. These precision values represent the typical performance from instruments
that are maintained properly, are operating in acceptable environmental conditions
and are using only recommended reagents and supplies.
Table 4.14 Typical Precision of Hemogram Parameters
Parameter CV
WBC (WOC) 1.9%
WBC (NOC) 1.9%
RBC 0.6%
HGB 0.4%
MCV 0.5%
RDW 2.3%
PLT 2.3%
MPV 1.8%