1-26
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Use or Function
System Components
Section 1
CELL-DYN Reagents
CELL-DYN 3200 Diluent/Sheath (List Number 03H79-01)
CELL-DYN Diluent/Sheath is formulated to meet the following requirements:
Act as the diluent for the RBCs, PLTs, and hemoglobin
Maintain the stable diluted cell volume of each red cell and platelet during
the count and sizing portion of the measurement cycle
Serve as a sheath fluid for the hydrodynamic focusing process
Serve as a rinsing agent for the fluidics system
Provide background counts equal to or less than:
WOC: 0.10 x K/µL
NOC: 0.10 x K/µL
RBC: 0.02 x M/µL
HGB: 0.10 x g/dL
PLT: 5.0 x K/µL
CELL-DYN 3200 CN-Free HGB/NOC Lyse (List Number 03H80-02)
CELL-DYN CN-Free HGB/NOC Lyse is formulated to meet the following
requirements:
Rapidly lyse the red blood cells and minimize the resultant stroma
Strip the white cell cytoplasm leaving the nuclear membrane intact so the
white cell nuclei can be enumerated
Convert hemoglobin to a stable chromagen complex that is measurable at
555 nm.
CELL-DYN 3200 WBC Lyse (List Number 03H78-01)
CELL-DYN WBC Lyse is formulated to meet the following requirements:
Act as the diluent for the WBCs
Osmotically lyse the red cells
Maintain the light scattering properties of the WBCs for the duration of the
measurement period
Provide sufficient wetting action to prevent accumulation of air bubbles in
the WBC flow system
Maintain a WOC background count equal to or less than 0.10 x K/µL
Act as a diluent for Reticulocytes