CELL-DYN
®
3200 System Operator’s Manual
6-37
9140181H—October 2001
Section 6 Calibration Procedures
Whole Blood Calibration Reference Values Worksheet
Date: Technologist:
NOTE: The WBC value obtained on the Reference Instrument should be used for
calibrating both the WOC and NOC parameters on the CELL-DYN 3200
Instrument.
Calibration Procedure — Open Mode
Follow the procedure below for calibrating the instrument in the Open Mode.
NOTE: Use the same Calibrator or Fresh Whole Blood samples for calibrating
Open and Closed Modes.
Determining New Calibration Factors
1. If necessary, go to the RUN menu and press [CHANGE SAMPLER] to select the
Open Mode.
2. In the MAIN MENU, press [CALIBRATION] to display the main CALIBRATION
menu.
Sample ID Run #
WBC
(WOC)
WBC
(NOC)
RBC HGB MCV PLT
1
2
1
2
1
2
1
2
1
2
Cumulative Mean