7-2
CELL-DYN
®
3200 System Operator’s Manual
9140181K—July 2002
Operational Precautions and Limitations
Overview
Section 7
Do not place reagent containers above the Analyzer.
The following space should be available to ensure proper ventilation:
Bench top space: approximately 6 linear feet to accommodate the
instrument, display station, and printer
Below the instrument: sufficient space for reagents and waste container
(if one is used)
Behind the instrument: 6 inches of space for proper ventilation
Above the instrument: 6 inches of space for proper ventilation
Left side of instrument: 12 inches of space in front of the fan for proper
ventilation
Allow adequate space around the instrument to perform necessary
maintenance procedures, to provide service access, and to allow the
instrument to be easily disconnected from its power source.
Care should be taken to prevent blocking of the air vents or fans on the sides
and the back of the instrument.
Before operating the instrument for the first time, verify that each reagent line
is connected to the appropriate inlet and reagent container. Refer to
Section 2: Installation Procedures and Special Requirements.
Make sure the waste line is connected to the appropriate outlet and routed to
a suitable waste container or drain. If the waste is routed to a waste container,
make sure the waste sensor is properly connected. If the waste is routed to a
drain, make sure a Dummy Plug is inserted in the Waste Sensor Connector.
The printer and Display Station can be placed on top of the instrument.