CELL-DYN
®
3200 System Operator’s Manual
6-53
9140181H—October 2001
Section 6 Calibration Procedures
1. In column 1, enter the new factor calculated in column 4 of the previous worksheet.
2. In column 2, enter the calibration factor that existed prior to running the current calibration procedure.
3. Subtract the current factor in column 2 from the new factor in column 1 and divide the result by the current
factor. Enter this result in column 3.
4. Multiply the result in column 3 by 100 and enter the result in column 5.
1. In column 1, enter the new Factor % Diff from column 5 of the previous worksheet (disregard the sign).
2. If the new Factor % Diff exceeds the limit in column 4, DO NOT CALIBRATE. Call the Abbott
Hematology Customer Support Center for assistance.
Worksheet 2
Open Mode Factor % Difference
(Calculate All Factors To Three Decimal Places)
(1)
New Open
Mode Factor
(2)
Current Open
Mode Factor
/
(3)
New Open
Mode Factor
(4)
x 100 =
(5)
Factor
% Diff
WOC
/ x 100 =
NOC
/ x 100 =
RBC
/ x 100 =
HGB
/ x 100 =
MCV
/ x 100 =
PLT
/ x 100 =
Worksheet 3
Open Mode Calibration Range Criteria
(1)
Factor%
Diff
(2)
Lower Limit
Cal Not Required
(3)
Calibration Range
Cal Required
(4)
Upper Limit
Do Not Cal
(5)
Cal? Y/N
WOC
<
1.5% >1.5% but <10% >10%
NOC
<
1.5% >1.5% but <10% >10%
RBC
<
1.0% >1.0% but <10% >10%
HGB
<
1.0% >1.0% but <10% >10%
MCV
<
1.0% >1.0% but <10% >10%
PLT
<
3.0% >3.0% but <15% >15%