CELL-DYN
®
3200 System Operator’s Manual
6-55
9140181H—October 2001
Section 6 Calibration Procedures
Worksheet 6
* Current factor as printed by the operator in the Preparing for Manual Mode to Mode Calibration section
within this procedure (Manual Mode to Mode Calibration).
** If factor exceeds limits, do not calibrate. Check all calculations and call the Abbott Hematology Customer
Support Center for assistance (at 1-800-CELL-DYN in the U.S.).
New Closed Mode Calibration Factors
(Calculate All Factors To Three Decimal Places)
Open Mode
Mean
/
Closed Mode
Mean
x
Current
Closed Mode
Cal Factor*
=
New Closed
Mode
Cal Factor
Range**
WOC
/ x = 0.700–1.300
NOC
/ x = 0.700–1.300
RBC
/ x = 0.800–1.200
HGB
/ x = 0.700–1.300
MCV
/ x = 0.700–1.300
PLT
/ x = 0.700–1.300
Open Mode Mean
Closed Mode Mean
X Current Closed Mode Cal Factor * = New Closed Mode Cal Factor