9-22
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Service and Maintenance
Monthly Maintenance Procedures
Section 9
Materials Required
1. Running water
2. DYN-A-WIPE lint-free pads (or other lint-free pads)
3. Small vacuum cleaner (optional)
Filter Cleaning Procedure
1. Perform the Daily Shutdown procedure to place the instrument in STANDBY.
(Refer to Daily Shutdown Procedure in Section 5: Operating Instructions.)
2. Turn the power switch to OFF.
3. Locate the Fan Filter on the rear panel. (Refer to Figure 9.7.)
4. Snap off the plastic frame which holds the filter in the mounting bracket.
5. Remove the filter and run a medium-pressure stream of warm water over
them, or clean the filter by vacuuming them.
6. Blot dry the filter with a lint-free pad.
7. Reinsert the cleaned filter into its frame and snap the frame back onto the
mounting bracket.
8. Turn the power switch to ON. After the instrument is initialized, press [RUN]
to prime the system and place it in the READY state.
9. Record this maintenance in your maintenance log.
10. If the filter appears unusually dirty when cleaned once a month, adjust the
cleaning cycle accordingly.