Section 10 Troubleshooting and Diagnostics
6. NOC:
• Follow the instructions given in step 5 for WBC, substituting the [NOC CNT]
key for the [WOC CNT] key for displaying and printing rate and graph data.
Troubleshooting Imprecise or Inaccurate Data
– – – - Probable Cause – – – – – – – - Corrective Action – – –
1. Improper Sample Mixing 1. Verify all samples are well-mixed prior to
analysis. (Refer to Section 5: Operating
Instructions; Subsection: Routine Operation.)
2. Dirty Syringe(s) 1. Inspect/Clean syringe as directed is Section 9:
Service and Maintenance; Subsection:
Nonscheduled Maintenance Procedures.
3. Leaking Syringe(s) 1. Inspect syringe for evidence of leaks.
2. Verify syringe connectors are secure.
3. Remove/Clean /Replace syringe as directed in
Section 9: Service and Maintenance;
Subsection: Nonscheduled Maintenance
Procedures.
4. Dirty Shear Valve 1. Clean shear valve as directed in Section 9: Service
and Maintenance; Subsection: Nonscheduled
Maintenance Procedures.
5. Dirty/Obstructed Y valve 1. Clean Y valve as directed in Section 9: Service
and Maintenance; Subsection: Nonscheduled
Maintenance Procedures.
6. Worn Sample Transfer Pump Tubing 1. Inspect/Replace sample transfer pump tubing as
directed in Section 9: Service and Maintenance;
Subsection: Nonscheduled Maintenance
Procedures.