CELL-DYN
®
3200 System Operator’s Manual
10-49
9140181H—October 2001
Section 10 Troubleshooting and Diagnostics
Message retransmission failure/BL
Communication from the Data Module to the Analyzer failed.
––– -Probable Cause–––– ––– -Corrective Action–––
1. A hardware failure or system error
has occurred.
1. Initialize the Analyzer by pressing
the [INITIALIZATION] key on the
second DIAGNOSTICS MENU
screen. If the message appears
again, reboot the System.
NOTE: If you wish to use the
Sample Loader, reset the
racks before initializing or
rebooting the System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)