CELL-DYN
®
3200 System Operator’s Manual
10-55
9140181H—October 2001
Section 10 Troubleshooting and Diagnostics
Download program timeout/BL
The allotted time for program downloading was exceeded.
– - Probable Cause – - Corrective Action
1. A failure in the Analyzer hardware,
possibly related to the HSSL cable,
card, or memory, has occurred.
1. Check the HSSL cable connections.
Then, initialize the Analyzer by
pressing the [INITIALIZATION] key
on the second DIAGNOSTICS MENU
screen. If the message appears
again, reboot the System.
NOTE: For the location of the
HSSL ports, refer to
Section 1: Use or
Function; Subsection:
Data Module
Components.
NOTE: If you wish to use the
Sample Loader, reset the
racks before initializing or
rebooting the System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)