Section 10 Troubleshooting and Diagnostics
Loader Fault 154: tube position bar code read failure/BL
A tube position bar code label (the second through tenth labels on the rack) could
not be read.
– - Probable Cause – - Corrective Action
1. The window on the Bar Code
Reader is dirty.
1. Remove the Loader Cover. Using a
DYN-A-WIPE pad or lint-free
wipe moistened with isopropyl
alcohol, lens cleaner or deionized
water, clean the window on the Bar
Code Reader. Dry the window on
the Bar Code Reader thoroughly.
Then, reinstall the Loader Cover,
reset the racks, and reset the Sample
Loader by pressing the following
keys in order:
[CLEAR FAULT],
[START LOADER],
[RESET LOADER].
NOTE: For the location of the Bar
Code Reader, refer to
Section 13: Sample
Loader; Subsection:
Sample Loader
Description.
2. The tube position bar code label is
scuffed or dirty.
2. Clean or replace the tube position
bar code label. Then, reset the racks
and reset the Sample Loader.
If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)