CELL-DYN
®
3200 System Operator’s Manual
11-17
9140181K—July 2002
Section 11 Quality Control
Running Controls
Abbott recommends using CELL-DYN Control materials for performing Quality
Control checks on the CELL-DYN 3200 System. Run a minimum of two levels of
control at the beginning of each eight hours of operation prior to running patient
samples. In addition, run controls after the following events:
After calibration (confirmatory step).
After Daily Start-Up procedures are completed.
After a reagent lot number change.
After maintenance, a service call, or component replacement.
After detection of an unusual trend or shift in patient results.
In accordance with the laboratory's Quality Control protocol.
According to regulatory requirements.
Controls should be run in both operating modes, (Open or Closed). Always do the
following:
If the System has been idle for 15 minutes or more, run a background
immediately prior to running a control specimen.
Follow the proper warming and mixing procedures previously described,
including those located on the control package insert.
Run control samples for each measured parameter in the same manner as
patient samples.
Verify that control results are within the laboratory's acceptable limits.
If the control results fall within acceptable limits, review the data for shifts or
trends, record the results, and begin to process patient samples.
If one or more result falls outside the laboratory's acceptable limits, review
Section 10: Troubleshooting and Diagnostics; Subsection:
Troubleshooting Guide. Perform suggested maintenance steps for the
parameter that is out-of-range, then try using another vial from the same lot.
If the problem persists, contact the Abbott Hematology Customer Support
Center.