CELL-DYN
®
3200 System Operator’s Manual
14-27
9140181H—October 2001
Section 14 Reticulocyte Package
Retic Data Log Codes
The following letters are displayed on the RETIC DATA LOG screen in the column
immediately preceding the date. (See the preceding figure.)
O Sample was run in the Open Mode.
K Flow Error or Fragile RBCs.
NOTE: Reticulocyte results are not suppressed for Fragile RBCs but are
suppressed for Flow Errors.
The RETIC DATA LOG screen contains the following soft keys:
EDIT ID (This key label displays only when the cursor is positioned
next to a patient record.)
DISPLAY SPECIMEN
FIND SPECIMEN
TRANSMIT DATA
PRINT DATA LOG
RETIC MAIN
Edit ID
The [EDIT ID] key on the RETIC DATA LOG screen is used to edit only the Specimen
ID. When the [EDIT ID] key is pressed, the cursor moves into the <SPECIMEN ID>
field and all soft key labels are blank. Each edit is saved by pressing the Enter key
on the keyboard.
NOTE: The [EDIT ID] key displays only when the cursor is positioned next to a
Reticulocyte Patient Record. It is not displayed for Reticulocyte
Background or Reticulocyte QC records.
The current date, time, and operator ID and the last Reticulocyte Sequence Number
that was used are displayed in the upper right-hand corner.