EasyManua.ls Logo

AGFA DR 800 - Page 112

AGFA DR 800
322 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1
2
4
3
1.
X-ray modality settings
2.
Generator settings for the static image
3.
Generator settings for fluoroscopy
4.
Automatic position
Figure 64: Examination overview
a) Check the X-ray modality settings.
Figure 65: X-ray modality controls on the software console
b) Check the exposure settings.
The fluo group contains settings for fluoroscopy and for the static
image.
Figure 66: Generator controls for fluoroscopy
Figure 67: Generator controls for the static image
3. Move the X-ray system to the right position.
a) Check if a correct automatic position is selected.
Figure 68: Positioning controls on the software console
b) Move to the selected automatic position.
112
| DR 800 | Basic Workflow
0392C EN 20210309 1049

Table of Contents

Related product manuals