Running Multiple-Vial Analyses
Recalibrating With Multiple Groups of Standards
Inj. Volume Default
Method METH2
8 Select the Calibration button [F1] to display the Calibration Settings screen for
sequence line 2 and choose Edit.
9 Enter the following information:
Calibr. vial range 4 to 6
#Inj 2
Inj. Volume Default
Calibr. method METH3
Recalibrate every 3 vials multi
Before On
After Off
10 Select the Done button [F6] to accept entries.
11 Select the Start button [F8] in the Sequence screen.
12 Press the Enter key to start the sequence.
The autosampler now analyzes:
three type A calibration standards in duplicate,
type A samples in vials 7 and 8,
three type A calibration standards in duplicate,
type A samples in vials 9 and 10,
three type A calibration standards in duplicate,
type A samples in vials 11 to 12,
three type A calibration standards in duplicate,
three type B calibration standards in duplicate,
type B samples in vials 13, 14 and 15,
three type B calibration standards in duplicate,
type B samples in vials 16,17 and 18,
three type B calibration standards in duplicate,
type B samples in vials 19, 20 and 21, and

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals