125
Running Multiple-Vial Analyses
Recalibrating With Multiple Groups of Standards
Inj. Volume Default
Method METH2
8 Select the Calibration button [F1] to display the Calibration Settings screen for
sequence line 2 and choose Edit.
9 Enter the following information:
Calibr. vial range 4 to 6
#Inj 2
Inj. Volume Default
Calibr. method METH3
Recalibrate every 3 vials multi
Before On
After Off
10 Select the Done button [F6] to accept entries.
11 Select the Start button [F8] in the Sequence screen.
12 Press the Enter key to start the sequence.
The autosampler now analyzes:
• three type A calibration standards in duplicate,
• type A samples in vials 7 and 8,
• three type A calibration standards in duplicate,
• type A samples in vials 9 and 10,
• three type A calibration standards in duplicate,
• type A samples in vials 11 to 12,
• three type A calibration standards in duplicate,
• three type B calibration standards in duplicate,
• type B samples in vials 13, 14 and 15,
• three type B calibration standards in duplicate,
• type B samples in vials 16,17 and 18,
• three type B calibration standards in duplicate,
• type B samples in vials 19, 20 and 21, and