Working with the Control Module
Automating Analyses
 • Line Number
Starting with 1 the sequence lines are automatically counted up.
 • Vial Range Information
Just like in the Analysis screen you can specify a range of vials together with the 
number of injections per vial.
 • Injection Volume
Although the injection volume is stored as a method parameter, a sequence has 
its own injection volume setting (overriding method information). If DEF is 
specified here the volume as set in the method is kept.
 • Method Name
Choose a method stored in the modules from the method list box (see “Working 
with Methods” on page 38). The method has to contain all the relevant 
parameters as well as timetable or injector program settings.
NOTE In the Sequence screen you create a program of several methods to be executed in a 
specified order. Editing of methods or their components (timetables, injector 
programs etc.) is not possible.
 • Wait Time
Specify a wait time that creates a gap between method loading and execution. 
This allows certain module parameters to stabilize before the next analysis is 
performed.
 • Calibration Settings
For each sequence line you can define calibration settings by pressing the 
Calibration button [F1] and choosing Edit/Delete from the pop up menu.
You can specify a range of calibration vials, number of injections, injection 
volume, calibration method and a wait time after method loading. You can also 
enter specific recalibration parameters, such as the recalibration interval and 
pattern. 
The Online Information System provides accurate information on recalibration 
options.