57
Working with the Control Module
Updating Firmware in the LC Modules
4 Select the file for the update (RESXXX.DLB), press the Execute button [F8] and
confirm with Yes.
In the screen you can monitor the status of the update procedure.
5 When the status bar has reached 100% press the Esc key.
The update of the resident firmware for the respective module has been completed.
Updating Operation Firmware
1 Prepare the system as described in “Preparing Firmware Update” on page 56.
2 Switch the module you want to update into resident mode as described in “Switching
Firmware Modes” on page 56. Also check if a resident firmware update is necessary.
3 Press System — Records and select the module you want to update.
4 Press the FW Update button [F5].
5 Choose the firmware file for the update (####-XXX.DLB), press the Execute button
[F8] and confirm with Yes.
NOTE Make sure you have selected the right firmware and module and the module is in the
resident mode as described above.
“Update started” will be displayed on the screen, then the progress of the firmware
update can be observed by the moving bar and the % display. If an “update failed,
wrong address” error message occurs, make sure that the module is not in the
resident mode (indicated by the flashing LED).
6 When the operation firmware update is finished press the Esc key to return to the
Records screen.
7 Press Transfer in the firmware update screen in order to return to normal operation
mode (see “Switching Firmware Modes” on page 56).
The control module is rebooted, the status lamp stops flashing and the Analysis
screen shows its normal appearance with this module.
8 Press System — Records to check that the correct firmware for the module was
loaded.
9 Reconnect the CAN cables or turn on the other modules.
When updating detector firmware, follow the update procedure with a detector
wavelength calibration.