59
Working with the Control Module
Coexecution with Agilent ChemStation
this specific entry field is disabled on the control module.
• If an analysis is running with the control module, the Agilent ChemStation should
not be turned on or rebooted.
• If the Agilent ChemStation starts an analysis, the control module is the slave/
monitor system.
• Parameter changes to a method will be identified on the other user interface as
modification.
• The control module and the Agilent ChemStation have a different method
handling (Agilent ChemStation method can have more information than the
method on the control module, e.g. additional DAD parameters that are only
accessible from the Agilent ChemStation). To have a method available on both
controllers proceed as follows:
If the method is on the Agilent ChemStation and should be saved on the control
module or PC card, load the method on Agilent ChemStation and then save the
method on the control module (or PC card) with Method — Save As.
If the method is on the control module or PC card and should be saved on the
Agilent ChemStation, first load method DEF_LC.M on the Agilent ChemStation
(to have no additional parameter in the format) and then load the required method
on control module. Then save the method on the Agilent ChemStation with the
same name.
WARNING A method that is available on the control module as protected method can be
modified by the Agilent ChemStation and then be saved on the control module
without any warning.