Agilent 1260 Infinity DAD and MWD User Manual 199
Maintenance
10
Exchanging a Lamp
5
Replace the front cover.
Next Steps:
6 Reset the lamp counter as described in the user interface
documentation (lamps with I.D. tag cannot be reset).
7 Turn the lamp on and give the lamp 10 minutes to warm
up.
8 Perform a “Wavelength Verification and Recalibration” on
page 183 or a “Holmium Oxide Test” on page 174 to check
the correct positioning of the UV-lamp.
9 Perform an “Intensity Test” on page 170.