202 Agilent 1260 Infinity DAD and MWD User Manual
10 Maintenance
Exchanging a Flow Cell
NOTE
The label attached to the flow cell provides information
on part number, path length, volume and maximum
pressure.
If you want to replace flow cell parts, “Maintenance of
Standard, Semi-Micro or Micro Flow Cell” on page 203
or “Maintenance of High Pressure Flow Cell” on
page 207.
6
Insert the flow cell while pressing the flow cell holder.
7 Insert the flow cell capillaries into the union holder (top is
inlet, bottom is outlet). Tighten the thumb screw and
Reconnect the waste tubing (bottom) to the union.
NOTE
To check for leaks, establish a flow and observe the
flow cell (outside of the cell compartment) and all
capillary connections.
Next Steps:
8 Perform a “Wavelength Verification and Recalibration” on page 183 or a “Holmium Oxide Test” on page 174 to check the
correct positioning of the flow cell.
9 Replace the front cover.