206 Agilent 1260 Infinity DAD and MWD User Manual
10 Maintenance
Maintenance of Standard, Semi-Micro or Micro Flow Cell
Next Steps:
9 Using a 4-mm hex key, tighten the window screw hand tight plus a quarter turn.
10 Reconnect the capillaries, see “Exchanging a Flow Cell” on page 200.
11 Perform a leak test.
12 Insert the flow cell.
13 Replace the front cover
14 Perform a “Wavelength Verification and Recalibration” on page 183 or a “Holmium Oxide Test” on page 174 to check the
correct positioning of the flow cell.