40 Agilent 1260 Infinity DAD and MWD User Manual
2 Site Requirements and Specifications
Performance Specifications
Communications Controller-area network (CAN), RS-232C, APG
Remote: ready, start, stop and shut-down
signals, LAN
Safety and
maintenance
Extensive diagnostics, error detection and
display (through control module and
ChemStation), leak detection, safe leak
handling, leak output signal for shutdown of
pumping system. Low voltages in major
maintenance areas.
GLP features RFID for electronics records of flow cell and UV
lamp conditions (path length, volume, product
number, serial number, test passed, usage)
Early maintenance feedback (EMF) for
continuous tracking of instrument usage in
terms of lamp burn time with user-setable limits
and feedback messages. Electronic records of
maintenance and errors. Verification of
wavelength accuracy with built-in holmium
oxide filter.
Housing All materials recyclable.
Others Electronic temperature control (ETC) for the
complete optical unit
Tab le 13 Performance Specifications G1315C/D and G1365C/D
Type Specification Comments