Non-volatile memory settings can be returned to their factory defaults.
Front Panel operation: return the settings to factory defaults under
the 11: PRESET command in the SYStem menu.
Remote operation: use the SYStem:PRESet command to return the
meter to the factory defaults.
Non-volatile memory settings and their factory defaults:
Setting Factory Default
Digital Filter
1
MEDium 50 rdgs
Digital Filter Precharge On
Low Power
Off
Low Volt
Off
LoV Limit
20 mV
Probe Type RTD
RTD Type
α = .00385
RTD R
0
100
Temperature Units
°C
Chart Out Off
Chart Span 1 mV
Chart Offset 0.00 V
Comma On
1
The digital filter is ON following a Front Panel PRESET.
The digital filter is OFF following a Remote Interface PRESET.
3
Chapter 3 Features and Functions
Defaults, Power-On and Reset States
113