• Independent vs. common settings: Some settings are independent to
the channel or function being used and some are common to multiple
channels or functions:
Voltage Resistance
Temperature
Channel 1 Channel 2
2-Wire Ω 4-Wire Ω
Null (On/Off) Null (On/Off) Null (On/Off) Null (On/Off)
Range Range Range
1
Range
Integration Time Integration Time Integration Time
Number of Digits Number of Digits Number of Digits
Math (Stats or Scale)
Digital Filter (On /Off)
Analog Filter (On/Off)
2
Trigger
Readings Hold
Readings Store
1
The meter may change the set resistance measurement range when voltage limited measurements
are enabled.
2
Analog filter available for voltage and thermocouple measurements on the 1 mV, 10 mV, and
100 mV ranges.
Chapter 3 Features and Functions
Defaults, Power-On and Reset States
114