Display Annunciators
*
Turns on during a measurement.
Adrs Meter is addressed to listen or talk over the HP-IB interface.
Rmt Meter is in remote mode (using remote interface).
Man Meter is using manual ranging (autorange is disabled).
Trig Meter is waiting for a single trigger or external trigger.
OC Off Offset compensation is turned off.
Null A null value is being used.
Stats Math statistics operations are being used.
Scale Math scaling operations are being used.
ERROR Hardware or remote interface command error(s) detected.
Shift "Shift" key has been pressed.
Ch1 Meter input is on Channel 1.
Ch2 Meter input is on Channel 2.
Ch1 - Ch2 Meter is indicating the difference between inputs on channel 1 and channel 2.
Ch1 / Ch2 Meter is indicating the ratio of the inputs on channel 1 and channel 2.
Ω Meter is measuring resistance (Ohms).
LP Meter is using low power Ohms.
2W Meter is using 2-wire Ohms (annunciator off indicates a 4-wire measurement).
LoV Meter is using voltage limited Ohms.
Hold Automatic reading hold is enabled.
Filt The analog and/or the digital filter is enabled.
Mem Turns on when reading memory is enabled.
To review the display annunciators, hold down the Shift key as you
turn on the meter.
4

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34420A and is the answer not in the manual?

Agilent Technologies 34420A Specifications

General IconGeneral
BrandAgilent Technologies
Model34420A
CategoryMultimeter
LanguageEnglish

Related product manuals