108 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
To adjust the multimode nebulizer
15 Adjust the electrospray needle position before reinstalling the nebulizer in
the spray chamber.
CAUTION
Take care when inserting the needle. The tapered end of the needle must pass through
the restrictions in the nebulizer shaft. The tip of the needle can be damaged if
excessive force is applied.
NOTE
Record this procedure in the Maintenance Logbook.