Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 59
Maintenance 2
To replace the electrospray nebulizer needle in the analytical or the reference nebulizer sprayer
Figure 12 Corroded needle in nebulizer
3 Loosen the locknut next to the zero-dead-volume (ZDV) union.
4 Remove the union from the nebulizer.
5 Loosen the locknut of the needle holder.
6 Unscrew the needle holder and pull it out of the nebulizer.
7 Slide the non-tapered end of the needle through the new needle holder from
the narrower side.
Figure 13 Sliding the non-tapered end of the needle through the new needle holder
8 Push a new ferrule, flat-side first, onto the needle.
9 Be sure the needle does not extend from the ferrule.
10 Reinstall the locknut and the union. Hand tighten the union.