Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 75
Maintenance 2
To clean the spray chamber weekly for the ESI with Agilent Jet Stream
Figure 22 Filling the spray chamber
5 Scrub the insulators and the interior of the spray chamber with a clean
cotton swab.
6 Empty the spray chamber.
7 Reinstall the spray chamber on the instrument.
8 Remove the spray shield.
9 Use abrasive paper to gently clean the end of the capillary cap.
WARNING
Some mobile phases are dangerous. Use the degree of caution appropriate for the
mobile phase being used.