76 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
To clean the spray chamber weekly for the ESI with Agilent Jet Stream
Figure 23 End of capillary cap
10 Dampen a clean cloth and wipe the end of the capillary cap.
11 Reinstall the spray shield.
12 Use abrasive paper to gently clean the spray shield.
13 Dampen a clean cloth and wipe the spray shield.
14 Rinse the area around the spray shield.
Figure 24 Rinsing the area around the spray shield
Capillary cap